• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. ARROW PERCUTANEOUS THROMBOLYTIC DEVICE SET: 5; CATHETER EMBOLECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. ARROW PERCUTANEOUS THROMBOLYTIC DEVICE SET: 5; CATHETER EMBOLECTOMY Back to Search Results
Model Number IPN035902
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 02/02/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
Customer reported "using trerotola device on an avf (arteriovenous fistula), we were working on the outflow, took the device out to get the clot off and when we took our image we noticed a perforation.Examined the device and the tip of the catheter was missing.We used a stent to cover up the perforation." it was reported the user had taken the device out to clean it and noticed the tip was off.A picture was taken using digital subtraction angiography and revealed a perforation.A stent was placed over the tip and perforation.The patient was reported to be well and did not have any additional complications.
 
Additional Manufacturer Narrative
Qn# (b)(4).The customer provided three photos for evaluation.The photos displayed the presumed pebax tip separated inside of a patient.A stent also appeared to be present, which confirms the customer report.The customer also returned one 5fr ptd catheter for evaluation.Visual examination revealed the pebax tip was fully separated from the ptd basket and not returned for evaluation.Microscopic examination confirmed the basket wires were fully welded.The returned ptd basket was able to be retracted and advanced from the ptd catheter.The returned ptd catheter was assembled with a lab inventory rotator.When the rotator button was depressed, the ptd catheter performed as expected.A device history record review was performed with no relevant findings.The instructions-for-use provided with this kit warns the user, "potential fatigue failure of ptd torque cable and fragmentation basket may occur with prolonged activation of ptd device.When using the ptd within the apex of a forearm loop graft, limit operation to 3 minutes or less to reduce the potential for basket failure.A rapid withdrawal rate of 1-2 cm/second is recommended when sharp radii are encountered (i.E., radius of loop graft, radii < 3 cm)." the customer report of a separated ptd tip was confirmed by complaint investigation of the returned sample.The ptd pebax tip was completely separated from the basket and was not returned for evaluation.A device history record review was performed with no relevant findings.A capa has previously been initiated to further investigate this issue.The root cause has been determined to be a manufacturing (molding) issue.Corrective actions have not yet been implemented.
 
Event or Problem Description
Customer reported "using trerotola device on an avf (arteriovenous fistula), we were working on the outflow, took the device out to get the clot off and when we took our image we noticed a perforation.Examined the device and the tip of the catheter was missing.We used a stent to cover up the perforation." it was reported the user had taken the device out to clean it and noticed the tip was off.A picture was taken using digital subtraction angiography and revealed a perforation.A stent was placed over the tip and perforation.The patient was reported to be well and did not have any additional complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW PERCUTANEOUS THROMBOLYTIC DEVICE SET: 5
Common Device Name
CATHETER EMBOLECTOMY
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
MDR Report Key11366862
Report Number9680794-2021-00084
Device Sequence Number934223
Product Code DXE
UDI-Device Identifier00801902013156
UDI-Public00801902013156
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K011056
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date08/31/2022
Device Model NumberIPN035902
Device Catalogue NumberPT-65509
Device Lot Number13F20J0771
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/19/2021
Initial Date Received by Manufacturer 02/08/2021
Supplement Date Received by Manufacturer03/01/2021
Initial Report FDA Received Date02/23/2021
Supplement Report FDA Received Date03/02/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-