| Model Number |
TRM-240 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
01/26/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture and patient's concern with the product.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side "rupture and exchange from textured to smooth due to concern with the product".Device remains implanted.
|
| |
|
Event or Problem Description
|
|
Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: analysis of the returned device identified: weight to the spec, broken shell, and brown particles in the shell.A visual and microscopic analysis was performed which identified: sharp broken on posterior and missing shell (0-25%).A dimension measurement in the shell was performed which identify the thickness within specification.Base on the device analysis the final assessment is: a sharp pattern on posterior of broken device assessed as an unidentified (tear) opening.Missing shell assessed as inconclusive.
|
| |
|