• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION PRISMAFLEX HF1400; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION PRISMAFLEX HF1400; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 107142
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/25/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that an external blood leak was observed during treatment with a prismaflex hf1400 set.It was observed that the blue ring connection on the return line continuously turned when tightened like the threads were stripped.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Additional Manufacturer Narrative
Additional information added h10.H10: the device was received for evaluation.A visual inspection and a leak test were performed and the reported condition was not verified.The cause could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX HF1400
Common Device Name
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11368335
Report Number8010182-2021-00062
Device Sequence Number1112744
Product Code KDI
UDI-Device Identifier07332414069315
UDI-Public(01)07332414069315
Combination Product (Y/N)Y
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K042938
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 03/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2022
Device Catalogue Number107142
Device Lot Number20H0602
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/10/2021
Initial Date Received by Manufacturer 01/28/2021
Supplement Date Received by Manufacturer03/16/2021
Initial Report FDA Received Date02/23/2021
Supplement Report FDA Received Date03/23/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-