| Model Number |
10605 |
| Medical Device Problem Codes |
Material Integrity Problem (2978); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/28/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: synergy ii us mr 3.00 x 28mm stent delivery system was returned for analysis.A visual examination of the stent found evidence of damage.The struts on the distal end lifted and pulled distally.The undamaged crimped stent outer diameter was measured and the result is within maximum crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual examination of the bumper tip showed signs of damage.A visual and tactile examination of the hypotube found no issues.A visual examination of the outer and inner lumen and mid-shaft section found no issues along the shaft polymer extrusion.Device tracked without issues on a test guidewire.No other issues were identified during the product analysis.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 04-feb-2021.It was reported via voluntary medwatch/maude report #: (b)(4) that guidewire loading difficulty and tip damage occurred.Post transcatheter aortic valve replacement, the patient presented to the emergency department with chest pain and non-st elevation myocardial infarction.When a 3.00 x 28mm synergy ii drug-eluting stent was opened, tip was damaged and unable to load into the wire.The procedure was completed with another of the same device and no patient complications were reported.However, returned device analysis revealed stent damage.
|
| |
|
Search Alerts/Recalls
|