Model Number 3241
Medical Device Problem Code Entrapment of Device (1212)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/11/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the burr became stuck on the wire.A 1.75mm rotalink plus and a rotawire and wireclip torquer were selected for use in an atherectomy procedure.The target lesion in the left anterior descending artery was severely calcified and located in moderately tortuous anatomy.During withdrawal while in dynaglide mode, the burr became stuck on the wire.The rotalink plus could not be pulled back, so both devices were removed together.The procedure was successfully completed.There were no patient complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTALINK PLUS
Common Device Name
CATHETER, CORONARY, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key11369350
Report Number2134265-2021-02065
Device Sequence Number3786371
Product Code MCX
UDI-Device Identifier08714729228370
UDI-Public08714729228370
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
P900056
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/11/2021
Device Model Number3241
Device Catalogue Number3241
Device Lot Number0024922505
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/11/2021
Initial Report FDA Received Date02/23/2021
Patient Sequence Number1
-
-