During investigation on-site performed by our field service engineer, the ventilator failed pressure transducer test during pre-use check.Visual inspection revealed moisture/water traces inside the o2 gas module.The conclusion was that the reported failed pressure transducer test was due to water traces inside the o2 gas module.The device logs were not received and no parts have been returned for investigation.Moisture/water in supply gases into a gas modules can cause failure which might lead to a stop of ventilation or high pressures.Alarms will be generated.According to the user's manual, maximum levels of water, (h2o < 7 g/m3) in the supplied gases to the ventilator must not be exceeded.The most probable source of moisture/water into a gas module is moisture from the hospital's gas supply system and/or external compressor.
|