Model Number 01.26.62MB
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code Pain (1994)
Date of Event 01/26/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on 02 february 2021: lot 081319: (b)(4) items manufactured and released on 14-oct-2008.Expiration date: 2013-08-31.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold.No other similar events have been reported since 2017.Clinical evaluation performed by medical affairs director: revision surgery performed 11 years after primary double-mobility total hip arthroplasty.No information concerning patient general health status is available.Radiographic images provided show suboptimal cup position: this may be the result of inadequate implant position during surgery or subsequent mobilization but this cannot be assessed not having immediate postoperative images; however, secondary cup migration is highly probable.The reason of this event cannot be determined, but there is no reason to suspect a faulty device.
 
Event or Problem Description
The patient came in reporting pain due to malpositioning of the cup.The cause of the malpositioned cup is unknown.The surgeon revised all the components with competitor hardware 10 years 11 months after the primary.The surgery was completed successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUP: VERSAFITCUP ACETABULAR SHELL 62
Common Device Name
ACETABULAR CUP
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ  6874
MDR Report Key11371451
Report Number3005180920-2021-00139
Device Sequence Number14462407
Product Code MEH
UDI-Device Identifier07630030808265
UDI-Public07630030808265
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K083116
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 02/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/31/2013
Device Model Number01.26.62MB
Device Catalogue Number01.26.62MB
Device Lot Number081319
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/26/2021
Initial Report FDA Received Date02/24/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-