• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. EQUINOXE; MDLR RV RMR 38MM PLT TIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. EQUINOXE; MDLR RV RMR 38MM PLT TIP Back to Search Results
Catalog Number 321-25-38
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/05/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Based on (b)(4), the broken device reported was likely the result of applying a bending moment to the reamer during use, which led to brittle fracture of the pilot tip feature.If action reported to fda under 21 usc 360i(f), list correction/removal reporting number: z-2663-2017, z-2664-2017, z-2665-2017, z-2666-2017, z-2668-2017, z-2668-2017.
 
Event or Problem Description
As reported the pilot tip on the end of the reamer broke of during the reaming portion of the procedure.The tip was found and we continued to ream with the next size up.This caused no delay, no harm to the patient, and we continued the process with no issues.The patient had no issues related to the procedure.Patient was last known to be in stable condition following the event.The device will return.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE
Common Device Name
MDLR RV RMR 38MM PLT TIP
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
MDR Report Key11371606
Report Number1038671-2021-00068
Device Sequence Number14137485
Product Code LXH
UDI-Device Identifier10885862087621
UDI-Public10885862087621
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
UNK
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 02/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number321-25-38
Device Lot Number113981018
Was Device Available for Evaluation? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/09/2021
Initial Report FDA Received Date02/24/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-