• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD 50/28MM BH DUAL MOBILITY INSERT; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD 50/28MM BH DUAL MOBILITY INSERT; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Model Number 74121650
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 01/28/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that a revision of a total right hip was performed.A smith & nephew unipolar head and insert were explanted.Smith & nephew synergy femoral stem and bhr cup remained in place.
 
Additional Manufacturer Narrative
H3, h6: it was reported that right hip revision surgery was performed.As of today, the implanted devices, all of which were used in treatment, and additional information have been requested for this complaint but have not become available.A review of the complaint history for the devices was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.No similar complaint were identified.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.Device history record review confirmed that all released parts met specifications applicable at the time of production.Review of the product's instructions for use found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as result of the reported event.The available medical documents were reviewed.The reported pseudotumor may be consistent with a reaction to metal debris.However, the source and the root cause cannot be determined with the available documentation.It cannot be concluded that the reported clinical findings are associated with the implant failure.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
50/28MM BH DUAL MOBILITY INSERT
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK  CV31 3HL
MDR Report Key11371667
Report Number3005975929-2021-00098
Device Sequence Number11698084
Product Code NXT
UDI-Device Identifier00885556577240
UDI-Public00885556577240
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P040033
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number74121650
Device Catalogue Number74121650
Device Lot NumberB2000463
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/28/2021
Supplement Date Received by Manufacturer04/29/2021
Initial Report FDA Received Date02/24/2021
Supplement Report FDA Received Date05/05/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-