• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH DELTA V-40 CERAMIC HEAD 36/+7,5; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH DELTA V-40 CERAMIC HEAD 36/+7,5; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS Back to Search Results
Model Number 6570-0-736
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/29/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Reported event: an event regarding crack/fracture involving a ceramic head was reported.The event was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: a review of the provided medical records by a clinical consultant stated the following comment: event description: as reported: "performed a left (mako) hip revision due to ceramic head fracture and gross trunnion failure.Original trident 2 cup stayed.No more information is available per doctor and implants will not be returned due to hospital policy." spoke to rep.The ceramic head cracked in half.Intraoperatively, wear and grooving of the trunnion of the stem was noted.It is not known if either device failure was the cause of the other.A stem, head and liner were revised.Rep confirmed there are no allegations against the revised liner.\ implant: delta v40 ceramic head 36mm +7.5mm, #6 accolade ii 1320 stem documents: 01/29/2021: stryker pi report.12/10/2019: op note, left primary daa mako hip, hand notes with gross trunnion failure, and head fx.Confirmation: the event cannot be confirmed.Root cause: the root cause cannot be ascertained without additional medical records and perhaps the retrieved implants.Product history review: review of the device history records indicates all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: a review of the provided medical records by a clinical consultant stated the following comment: the event cannot be confirmed [.] the root cause cannot be ascertained without additional medical records and perhaps the retrieved implants.The exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, pathology reports, pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
As reported: "performed a left (mako) hip revision due to ceramic head fracture and gross trunnion failure.Original trident 2 cup stayed.No more information is available per doctor and implants will not be returned due to hospital policy." spoke to rep.The ceramic head cracked in half.Intraoperatively, wear and grooving of the trunnion of the stem was noted.It is not known if either device failure was the cause of the other.A stem, head and liner were revised.Rep confirmed there are no allegations against the revised liner.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DELTA V-40 CERAMIC HEAD 36/+7,5
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key11371835
Report Number0002249697-2021-00327
Device Sequence Number3773734
Product Code LZO
UDI-Device Identifier04546540608581
UDI-Public04546540608581
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K052718
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2021
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2024
Device Model Number6570-0-736
Device Catalogue Number6570-0-736
Device Lot Number75358301
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/29/2021
Initial Report FDA Received Date02/24/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age48 YR
Patient Weight99
-
-