DATASCOPE CORP. - FAIRFIELD TRANS-RAY PLUS 7.5 FR. 40CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Catalog Number 0684-00-0605
Device Problems Difficult to Insert (1316); Unraveled Material (1664)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/07/2021
Event Type  malfunction  
Manufacturer Narrative
Event site postal code: (b)(6).The device was not returned and could not be evaluated.It was discarded by the user.We are unable to confirm the reported event.If new information becomes available, a supplemental report will be submitted.Complaint record id #: (b)(4).
 
Event Description
It was reported that during insertion of the intra-aortic balloon (iab), the iab was unable to pass through the sheath due to the balloon membrane becoming unfurled.A new iab was used instead to continue therapy.There was no patient harm or adverse event reported.
 
Manufacturer Narrative
The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.The failure mode is addressed in the risk file and is operating within its risk profile.The ifu addresses the reported failure.There were no ncmrs identified which could cause or contribute to the reported failure.The investigation does not indicate that the device was inadvertently released as non-conforming or an adulterated product or was a counterfeit.The complaint history review did not identify an adverse trend (increase in number of complaints over past three (3) months).Based on the rational provided above, no escalation to the capa process is required.Reference complaint (b)(4).
 
Event Description
It was reported that during insertion of the intra-aortic balloon (iab), the iab was unable to pass through the sheath due to the balloon membrane becoming unfurled.A new iab was used instead to continue therapy.There was no patient harm or adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRANS-RAY PLUS 7.5 FR. 40CC IAB
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ 07004
MDR Report Key11371897
MDR Text Key233262134
Report Number2248146-2021-00125
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 04/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/24/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/19/2023
Device Catalogue Number0684-00-0605
Device Lot Number3000117326
Was Device Available for Evaluation? No
Date Manufacturer Received04/16/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-