DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-3023-55 |
Device Problem
Connection Problem (2900)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/01/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge field service engineer (fse) was dispatched to investigate.The fse ran the unit for approximately an hour with the defective ecg cable, as well as the replacement ecg cable without any issues.The fse was unable to reproduce the reported issue.Additionally, the fse did not find any abnormalities.The fse then performed functional and safety checks to meet factory specifications.The iabp was then released to the customer and cleared for clinical service.
|
|
Event Description
|
It was reported that during use on a patient, the cs300 intra-aortic balloon pump (iabp) ecg cable was faulty, whereby the unit showed flat line.It was also reported that the end user replaced the ecg cable with no issues and continued treatment, thereafter.No patient harm, serious injury or adverse event was reported.
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
Updated fields: b4, b6, b7, d1 (version or model #), d10, g3, g6, g7, h2, h4, h6, h10, h11 corrected fields: d1 (brand name).
|
|