DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number N/A |
Device Problems
Failure to Zero (1683); Failure to Calibrate (2440)
|
Patient Problems
No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/09/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge service territory manager (stm) was dispatched to investigate the calibration flashing and unable to zero pressure.The stm was unable to reproduce the issue.The stm reported that when testing with balloon and trainer, the stm was able to zero with out any problem.Then the stm ran calibration and functional tests and everything passed.The unit passes all tests and calibrations according to manufacturing specifications.The iabp was returned to the customer for clinical use.
|
|
Event Description
|
It was reported that before use, the cardiosave intra-aortic balloon pump (iabp) was not able to zero and calibration was flashing.There was no patient involvement, and no adverse event reported.
|
|
Manufacturer Narrative
|
Analysis of production: (3331/213) the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the dhr/lhr review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis: (4109/213) the review of the historical data indicates that no other similar complaint was reported for the same lot/serial number and reported failure mode.Trend analysis: (4110/213) the overall 24 month product complaint trend data for the period (mar 2019 through feb 2021) was reviewed.There were no triggers identified for the review period.
|
|