• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER TRAUMA KIEL ADJUSTING DEVICE FEMUR ANTEGRADE IMN INSTRUMENTS; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER TRAUMA KIEL ADJUSTING DEVICE FEMUR ANTEGRADE IMN INSTRUMENTS; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 2353-3105
Medical Device Problem Code Difficult or Delayed Separation (4044)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/29/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Upon completion of investigation, additional information will be provided in a supplemental report.
 
Event or Problem Description
As reported: "it was difficult to remove the drill sleeve that had been inserted into the adjusting device of the distal target device.The lock mechanism was stiff, blood was clogged and the lock was stiff.Please check if there are any defects in the adjusting device itself." additional information: according to the report, the blood was clots and lock mechanism was hard.No medical intervention measures were needed.
 
Additional Manufacturer Narrative
The reported event, instrument - assembling / disassembling impaired during implantation, was not confirmed.The devices were returned in cleaned condition and could be assembled / disassembled as intended.Function of clamping / releasing feature (for sleeves) and height adjustment was fully given.The sleeves could be inserted and removed easily.No deviations were found during review of the manufacturing and inspection documents (dhr).The returned adjusting device was documented as faultless prior to distribution.The risk management file review showed that the complained event was addressed adequately; no threshold exceedances found.No nc/capa record was filed regarding the item in question with this failure mode.Also, no further events regarding dried blood was reported.With available information no product deficiency was determined.The event was regarded as user related.In case further information becomes available we reserve the right to reopen the investigation and to re-assess the root cause.
 
Event or Problem Description
As reported: "it was difficult to remove the drill sleeve that had been inserted into the adjusting device of the distal target device.The lock mechanism was stiff, blood was clogged and the lock was stiff.Please check if there are any defects in the adjusting device itself." additional information: according to the report, the blood was clots and lock mechanism was hard.No medical intervention measures were needed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADJUSTING DEVICE FEMUR ANTEGRADE IMN INSTRUMENTS
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
MDR Report Key11373086
Report Number0009610622-2021-00292
Device Sequence Number11698068
Product Code HSB
UDI-Device Identifier07613327357035
UDI-Public07613327357035
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K191271
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2353-3105
Device Catalogue Number23533105
Device Lot NumberKME914877
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/25/2021
Initial Date Received by Manufacturer 01/29/2021
Supplement Date Received by Manufacturer04/14/2021
Initial Report FDA Received Date02/24/2021
Supplement Report FDA Received Date05/06/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-