| Model Number |
400CE |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/27/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
During surgery the machine heated the water to 42°c, but when the perfusionist turned on the pump, the heater turned off and it never turn on again.The temperature dropped to 30°c but the setpoint was in 42°c.Later, they checked the machine and when turned on the pump in the heater mode, the machine did not heat the water.But if they stop the pump, the heater started to heat.No adverse event reported.
|
| |
|
Event or Problem Description
|
|
During surgery the machine heated the water to 42°c, but when the perfusionist turned on the pump, the heater turned off and it never turn on again.The temperature dropped to 30°c but the setpoint was in 42°c.Later, they checked the machine and when turned on the pump in the heater mode, the machine did not heat the water.But if they stop the pump, the heater started to heat.No adverse event reported.
|
| |
|
Search Alerts/Recalls
|
|
|