| Catalog Number |
315-03-01 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/03/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Pending evaluation.
|
| |
|
Event or Problem Description
|
|
As reported, during a initial shoulder procedure on a (b)(6) female, the surgeon reduced shoulder to trial selected implants, utilized skid to dislocate trials, and skid snapped.Patient was last known to be in stable condition following the event.The device will return for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
Section h10: (h3) based on previous similar complaint investigations, the broken glenoid skid reported was likely the result of applying a bending moment to one end of the instrument which led to fracture of the device.
|
| |
|