• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER LITHIUM HEPARINN (LH) 37 USP UNITS BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER LITHIUM HEPARINN (LH) 37 USP UNITS BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Model Number 366664
Medical Device Problem Codes Short Fill (1575); Volume Accuracy Problem (1675)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/09/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.(b)(4).
 
Event or Problem Description
It was reported the bd vacutainer® lithium heparinn (lh) 37 usp units blood collection tubes the client experienced underfill or low draw of a tube with blood.The following information was provided by the initial reporter.The customer stated: "during post market surveillance activity phone call, customer complained of have frequent sample underfill on bd vacutainer ®sodium fluoride tube, bd vacutainer ®lithium heparin tube, and bd vacutainer ®edta tube.She mentioned that the fluoride tube has 2021-03-31 expiration date, the lithium heparin tube has june 2021 expiration, but did not provide reference numbers or lot number for any of the tubes.".
 
Event or Problem Description
It was reported the bd vacutainer® lithium heparinn (lh) 37 usp units blood collection tubes the client experienced underfill or low draw of a tube with blood.The following information was provided by the initial reporter.The customer stated: "during post market surveillance activity phone call, customer complained of have frequent sample underfill on bd vacutainer ®sodium fluoride tube, bd vacutainer ®lithium heparin tube, and bd vacutainer ®edta tube.She mentioned that the fluoride tube has 2021-03-31 expiration date, the lithium heparin tube has june 2021 expiration, but did not provide reference numbers or lot number for any of the tubes.".
 
Additional Manufacturer Narrative
H.6.Investigation: bd had not received samples or photos for investigation.Therefore, 10 retention samples from bd inventory were evaluated by functional testing and no issues were observed relating to underfill as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.Bd was not able to identify a root cause for the indicated failure mode.H3 other text : see h.10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER LITHIUM HEPARINN (LH) 37 USP UNITS BLOOD COLLECTION TUBES
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
MDR Report Key11376196
Report Number1917413-2021-00141
Device Sequence Number19842949
Product Code JKA
UDI-Device Identifier50382903666640
UDI-Public50382903666640
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K945952
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date03/31/2021
Device Model Number366664
Device Catalogue Number366664
Device Lot Number9315399
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/09/2021
Supplement Date Received by Manufacturer03/08/2021
Initial Report FDA Received Date02/24/2021
Supplement Report FDA Received Date03/11/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-