Model Number 799-EMVP-04-01 |
Device Problem
Communication or Transmission Problem (2896)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
|
Event Description
|
Complainant alleged that during a routine shift check by a clinician, the device displayed an "internal comm failure -1474" message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
|
Manufacturer Narrative
|
The device was returned to a zoll approved service provider for evaluation.The customer's report was observed during initial testing.However, the report could not be duplicated.The device is expected to be returned to zoll medical corporation.Analysis of reports of this type has not identified an increase in trend.
|
|
Manufacturer Narrative
|
The device was returned to zoll medical corporation; the customer's report was duplicated and attributed to fractured solder on an integrated circuit on the smart pneumatic module.The smart pneumatic module was replaced to remedy the report.The device was recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
|