| Model Number |
PCB00 |
| Medical Device Problem Codes |
Break (1069); Crack (1135)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
01/06/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If implanted, give date: not applicable, as lens was not implanted.If explanted, give date: not applicable, as lens was not implanted.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the inserter of the preloaded device bent over the intraocular lens (iol) instead of pushing it out.There was no patient contact.The event was observed when inserting into the eye.No further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction and additional information: section b5: additional: subsequent follow up information received from the surgery center confirms that the issue inserter bent over the lens was associated with a different serial number (sn(b)(6) and the actual issue associated with the serial number captured in the report medwatch 2648035-2021-07315 was torn cartridge.There was no patient contact.Section h6: additional: medical device problem code: 1135 - crack upon reviewing the complaint folder, it has been determined that the original issue with the preloaded intraocular lens (iol) was torn cartridge and the issue occurred prior to any patient contact.Based on our reportable criteria, this report is no longer a reportable event.Therefore, an emdr is required to state that no further follow ups will be sent out under medwatch 2648035-2021-07315.G8: correction: in review of medwatch 2648035-2021-07315 and per new information received, this event has been assessed not reportable.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|