Model Number 37303
Medical Device Problem Code Nonstandard Device (1420)
Health Effect - Clinical Codes Skin Tears (2516); Localized Skin Lesion (4542)
Date of Event 02/03/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the male external catheters had a good grip which made it difficult to remove.It was reported that it caused skin lesions.
 
Additional Manufacturer Narrative
The reported event was inconclusive as no sample returned for evaluation.A potential root cause for this failure mode could be due to viscometer failure or mechanical failure.It was unknown whether the device had met specifications.The product used for the treatment but it was unknown whether the product had caused the reported failure.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "indication the self adhering male external catheter is designed for the management of male urinary incontinence.Contraindication do not use on irritated or compromised skin.Precaution do not use if allergic reaction occurs.For good hygiene, change catheter daily.Use of a single device for longer periods than 24 hours may increase the risk of complications.Directions: to apply 1) wash penis with mild soap and warm water.Dry thoroughly.2) trim pubic hair if necessary.3) open package at perforation.4) to remove plastic insert, squeeze catheter at the top of the white cone and pull to release.5) unroll self adhering catheter over penis.6) gently squeeze the catheter to properly seal adhesive to the skin.7) connect to collection bag.Important: wear time may be significantly reduced if adhesive is not properly sealed to the skin.Directions: to remove gently roll catheter off the penis.Note: if necessary, apply a warm wet compress (such as a wet washcloth) around the catheter to help loosen the adhesive." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Event or Problem Description
It was reported that the male external catheters had a good grip which made the catheter difficult to remove and caused the skin lesions.No medical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPIRIT HYDROCOLLOID ADHESIVE SHEATH
Common Device Name
MALE EXTERNAL CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
MDR Report Key11378732
Report Number1018233-2021-00808
Device Sequence Number17094767
Product Code NNX
UDI-Device Identifier00801741071195
UDI-Public(01)00801741071195
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2024
Device Model Number37303
Device Catalogue Number37303
Device Lot NumberJUDX0241
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/03/2021
Supplement Date Received by Manufacturer03/19/2021
Initial Report FDA Received Date02/25/2021
Supplement Report FDA Received Date04/12/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-