Model Number N/A |
Device Problem
Incorrect Measurement (1383)
|
Patient Problems
No Information (3190); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/05/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Customer had not re-engaged console into cart.Re-engaged after instruction and issue resolved.Additional information is being requested with regard to the repair and status of the iabp.A supplemental report will be submitted when this information is provided to us.No service requested.
|
|
Event Description
|
It was reported that the cardiosave intra-aortic balloon pump (iabp) analog helium gauge showed empty when the helium gauge on the screen showed half full.It is unknown the circumstances under which the event occurred.It is also unknown if there was a patient involvement.However, there was no adverse event reported.
|
|
Event Description
|
It was reported that the cardiosave intra-aortic balloon pump (iabp) analog helium gauge showed empty when the helium gauge on the screen showed half full.It is unknown the circumstances under which the event occurred.It is also unknown if there was a patient involvement.However, there was no adverse event reported.
|
|