C.R. BARD, INC. (BASD) -3006260740 DOT PPICC SOLO PROVENA 4F DL 3CG; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number CK000409A
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/27/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, complaint and batch history, applicable previous investigation(s), labeling, applicable manufacturing records, sample analysis and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a catheter kink is confirmed but the exact caused remains unknown.Two radiographic images of a catheter within a patient were provided for evaluation.The first photo shows a powerpicc catheter inserted within a patients arm.The catheter appears to twist and kink at two locations on the proximal end of the catheter.The second image appears to show the same catheter; however, only one kink appears to be present.This may suggest manipulation of the catheter during use.Based on the description of the reported event, possible contributing factors include insertion method, securement of catheter, and excessive manipulation of the catheter during use.Since the radiographic image appears to show evidence of a kinked catheter, the complaint is confirmed but the exact cause remains unknown.A lot history review (lhr) review is not possible, as no manufacturing lot number has been provided by the complainant.
 
Event or Problem Description
It was reported "placed on 1/26 ck000409a 4 french dual lumen powerpicc provena solo, the next day no blood return, xray taken of arm and kink found.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DOT PPICC SOLO PROVENA 4F DL 3CG
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
MDR Report Key11379883
Report Number3006260740-2021-00490
Device Sequence Number14423682
Product Code LJS
UDI-Device Identifier00801741215070
UDI-Public(01)00801741215070
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K180548
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCK000409A
Device Catalogue NumberCK000409A
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/02/2021
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/27/2021
Initial Report FDA Received Date02/25/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-