| Model Number |
N/A |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/02/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Evaluation of the returned device found that the operates as intended.The device exhibited usage mark however dimensional analysis found the device to be within specification.Functional dimensions were performed and found to be acceptable.The evaluation of device found that the device operated within specification and hence no issue with the device was seen.The complaint cannot be confirmed.No corrective actions, preventive actions, or field actions resulted after investigation of this event.
|
| |
|
Event or Problem Description
|
|
It was reported the femoral inserter/extractor handle is broken.The spring isn't working anymore.No impact on patient.
|
| |
|