Additional Manufacturer Narrative
|
|
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
Device was returned with packaging damage and sterility barrier was compromised.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected: b4, b5, g4, h2, h3, h4, h6.Event was confirmed by visual examination of the returned product.Visual evaluation identified that the outer carton and outer sterile cavity had been damaged.Review of the device history records identified no deviations or anomalies during manufacturing.The product was likely conforming when it left zimmer biomet control.The root cause has been attributed to transit damage.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No additional information at the time of this report.
|
| |
|