| Model Number |
4078 |
| Medical Device Problem Code |
Sensing Intermittently (1558)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/04/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.
|
| |
|
Event or Problem Description
|
|
The customer reported intermittent spo2 readings.No patient impact or consequences were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional manufacturing narrative: the returned device was evaluated.External visual inspection found contamination on and around the pogo pins.The device failed continuity tests due to contamination causing pogo pin #3 to stick in the barrel.As a result, the cable has an open connection on the yellow conductor.The cable was malfunctioning with no reading.The connected monitor audibly and visually alarmed.The sensor led does not illuminate.
|
| |
|
Event or Problem Description
|
|
The customer reported intermittent spo2 readings.No patient impact or consequences were reported.
|
| |
|