| Model Number |
IPN049808 |
| Medical Device Problem Code |
Failure to Deliver (2338)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/28/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed and no relevant findings were identified.Without the device to evaluate, the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
|
| |
|
Event or Problem Description
|
|
It was reported "the doctor found no mist during clinical inspection before using on patient".No patient involvement reported.
|
| |
|
Event or Problem Description
|
|
It was reported "the doctor found no mist during clinical inspection before using on patient".No patient involvement reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4), the sample was received for evaluation.A visual exam was performed and a foreign particle was found in the nebulizer jar.A device history record review was performed and no relevant findings were identified.Based on the visual exam the complaint was confirmed.A non-conformance was opened to address this issue.
|
| |
|