• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HUDSON MICRO MIST NEBULIZER W/ELONG; NEBULIZER (DIRECT PATIENT INTE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HUDSON MICRO MIST NEBULIZER W/ELONG; NEBULIZER (DIRECT PATIENT INTE Back to Search Results
Model Number IPN049808
Medical Device Problem Code Failure to Deliver (2338)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/28/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed and no relevant findings were identified.Without the device to evaluate, the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
 
Event or Problem Description
It was reported "the doctor found no mist during clinical inspection before using on patient".No patient involvement reported.
 
Event or Problem Description
It was reported "the doctor found no mist during clinical inspection before using on patient".No patient involvement reported.
 
Additional Manufacturer Narrative
(b)(4), the sample was received for evaluation.A visual exam was performed and a foreign particle was found in the nebulizer jar.A device history record review was performed and no relevant findings were identified.Based on the visual exam the complaint was confirmed.A non-conformance was opened to address this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON MICRO MIST NEBULIZER W/ELONG
Common Device Name
NEBULIZER (DIRECT PATIENT INTE
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
MDR Report Key11382219
Report Number3004365956-2021-00068
Device Sequence Number14475305
Product Code CAF
UDI-Device Identifier14026704645937
UDI-Public14026704645937
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K930525
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/11/2024
Device Model NumberIPN049808
Device Catalogue Number1886
Device Lot Number74B1900530
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/05/2021
Initial Date Received by Manufacturer 02/07/2021
Supplement Date Received by Manufacturer04/07/2021
Initial Report FDA Received Date02/25/2021
Supplement Report FDA Received Date04/07/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-