SMITH & NEPHEW, INC. OSTEORAPTOR 2.3 W 1UB II BLUE; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
| Model Number |
72201991 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/26/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the osteoraptor's tip broke during insertion in an unknown procedure.The broken pieces were removed from the patient by suction, and the procedure was completed with a smith and nephew back up device.No delays or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A review of risk management files found that the reported failure was documented appropriately.A review of the instructions for use found the following warnings and precautions related to the reported failure: breakage of the suture anchor can occur if the insertion site is not prepared with the recommended drill prior to implantation.Ensure the anchor placement is aligned with the drilled hole.Proper alignment is essential for successful repair.¿ use of excessive force during insertion can cause failure of the suture anchor or insertion device.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|
| |
|
Search Alerts/Recalls
|
|
|