MAUDE Adverse Event Report: SMITH & NEPHEW, INC. OSTEORAPTOR 2.3 W 1UB II BLUE; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. OSTEORAPTOR 2.3 W 1UB II BLUE; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE Back to Search Results
Model Number 72201991
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/26/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event or Problem Description
It was reported that the osteoraptor's tip broke during insertion in an unknown procedure.The broken pieces were removed from the patient by suction, and the procedure was completed with a smith and nephew back up device.No delays or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
The reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A review of risk management files found that the reported failure was documented appropriately.A review of the instructions for use found the following warnings and precautions related to the reported failure: breakage of the suture anchor can occur if the insertion site is not prepared with the recommended drill prior to implantation.Ensure the anchor placement is aligned with the drilled hole.Proper alignment is essential for successful repair.¿ use of excessive force during insertion can cause failure of the suture anchor or insertion device.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSTEORAPTOR 2.3 W 1UB II BLUE
Common Device Name
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
MDR Report Key11382278
Report Number1219602-2021-00411
Device Sequence Number14296061
Product Code MAI
UDI-Device Identifier03596010615527
UDI-Public03596010615527
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K151105
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/08/2025
Device Model Number72201991
Device Catalogue Number72201991
Device Lot Number2061380
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/05/2021
Supplement Date Received by Manufacturer03/30/2021
Initial Report FDA Received Date02/25/2021
Supplement Report FDA Received Date04/23/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-