Updated fields: b4, g4, g7, h2, h4, h6 (health effec-clinical codes), (type of investigation), (component codes), (health effect-impact codes), h10.Analysis of production: (3331/213) the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported serial number.Based on the dhr/lhr review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis: (4109/213) the review of the historical data indicates that no other similar complaint was reported for the same serial number and reported failure mode.Trend analysis: (4110/213) the overall 12 month product complaint trend data for the period (feb 2020 through jan 2021) was reviewed.There were no triggers identified for the review period.
|