• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CHRISTOPH MIETHKE GMBH & CO KG PROGAV 2.0 SYS PED.W/SA20 A.BURRHOLE RES.; HYDROCEPHALUS MANAGEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CHRISTOPH MIETHKE GMBH & CO KG PROGAV 2.0 SYS PED.W/SA20 A.BURRHOLE RES.; HYDROCEPHALUS MANAGEMENT Back to Search Results
Model Number FX448T
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 01/06/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation on going.The release for investigation was received on 02/22/2021.Investigation result will be provided in a supplemental report.
 
Event or Problem Description
It was reported that there was a problem with a progav 2.0 valve.The surgeon explanted the valve, because the valve is not further adjustable to the pressure setting.Patient informations: age y: (b)(6), height cm: 85, weight kg: (b)(6), gender: unknown, implantation: (b)(6) 2019 , explantation: (b)(6) 2021.
 
Event or Problem Description
Investigation is done on 01/03/2021.
 
Additional Manufacturer Narrative
Investigation: visual inspection: no significant deformations or damage of the valves were detected during the visual inspection.Permeability test: a permeability test has shown that both valves are permeable.Adjustment test: the progav 2.0 was tested and is adjustable to all specified pressures.Braking force and brake function test: the brake functionality test has shown that the brake function is fully operational and the braking force is within the given tolerances.Results: first, we performed a visual inspection of the progav 2.0 shunt system.No significant deformations or damage of the valves were detected during the visual inspection.Next, we tested the permeability, adjustability, as well as the brake functionality and brake force.The valve operates as expected and met all specifications.Finally, we have dismantled the valves.Inside both valves we have found slight build-up of substances (likely protein).Based on our investigation, we are unable to substantiate the claim of non-adjustability.At the time of our investigation, the valve was able to be adjusted to all specified settings.However, it is possible that the deposits observed inside the valve could have caused the malfunction in the past.As described in scientific literature, the problem encountered is one of the known, inevitable risks of hc-therapy by shunt implants.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGAV 2.0 SYS PED.W/SA20 A.BURRHOLE RES.
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
MDR Report Key11385777
Report Number3004721439-2021-00031
Device Sequence Number3283112
Product Code JXG
UDI-Device Identifier04041906137748
UDI-Public4041906137748
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K161853
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/15/2023
Device Model NumberFX448T
Device Catalogue NumberFX448T
Device Lot Number20037147
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/02/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/02/2021
Supplement Date Received by Manufacturer03/01/2021
Initial Report FDA Received Date02/26/2021
Supplement Report FDA Received Date03/25/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age1 YR
Patient Weight11
-
-