Model Number PCB00
Medical Device Problem Codes Difficult to Advance (2920); Mechanical Jam (2983); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/07/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If implanted, give date: not applicable, as lens was not implanted.If explanted, give date: not applicable, as lens was not implanted.Device evaluation: the complaint product was returned in its original packaging.Upon evaluation all the components were correctly engaged.No assembly error and/or defect related to the manufacturing process was identified.Plunger was in a partially advanced position.Traces of lubricating material can be observed in the device.The lens was out of the tip, partially stuck at the cartridge tip, and it looks torn.Due to the conditions of the sample returned the reported issue could not be associated to manufacturing issue.The stuck in cartridge was verified, but a product quality deficiency could not be determined.Manufacturing record review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.There are no discrepancies found during the mrr (manufacturing record review).A search revealed that no additional complaint was received from this production order.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the preloaded intraocular lens (iol) was stuck in cartridge.The lens was partially inserted.The event was observed while inserting lens into eye.There was no incision enlargement.No further information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS ITEC PRELOADED 1-PIECE IOL
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key11386928
Report Number2648035-2021-00034
Device Sequence Number17184286
Product Code HQL
UDI-Device Identifier05050474558243
UDI-Public(01)05050474558243(17)220314
Combination Product (Y/N)Y
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,health profession
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/14/2022
Device Model NumberPCB00
Device Catalogue NumberPCB0000190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/22/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/11/2021
Initial Report FDA Received Date02/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-