PORTEX; TRACHEOSTOMY Back to Search Results
Model Number 101/860/085CZ
Medical Device Problem Codes Fluid/Blood Leak (1250); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Event or Problem Description
Information received a smiths medical tracheostomy|pvc - portex tubes bluselect malfunctioned.The customer confirmed that the product worked properly in a pre-use check.However, during the use of the product, the customer noticed the inflation line got detached from the pilot balloon.No patient injury.
 
Additional Manufacturer Narrative
Other, other text: one used decontaminated sample 10009163-005 8.5mm deht blu suctionaid sub-assy was received in plastic bag without its original packaging (see photo of sample).Under visual inspection we noticed that pilot balloon was detached from inflation line.10001429-005 infl.Line s/assy soft-seal 8.5mm is assembled in smiths medical tijuana therefore secondary investigation object in-0099123 was created and complaint sample was resent to this site.
 
Event or Problem Description
Investigation completed and summary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX
Common Device Name
TRACHEOSTOMY
MDR Report Key11387139
Report Number3012307300-2021-01599
Device Sequence Number11606120
Product Code BTO
UDI-Device Identifier15019517076165
UDI-Public15019517076165
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K173384
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number101/860/085CZ
Device Catalogue Number101/860/085CZ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/10/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 02/05/2021
Supplement Date Received by Manufacturer07/21/2021
Initial Report FDA Received Date02/26/2021
Supplement Report FDA Received Date08/19/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
-
-