| Model Number |
1218-28-050 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Pain (1994); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
07/31/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Implantation of a hip-tep right side because of hip-head necrosis on (b)(6) 2007.In (b)(6) 2020 a conspicuous sound was heard in the hip, no trauma, since (b)(6) 2020 strong pain.Revision on (b)(6) 2020 because of dislocation of inlay.Contact of ceramic hip head to cup.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.No evidence was found indicating product error was a contributing factor.The need for corrective action was not indicated.Depuy considers the investigation closed.Should additional info be received, the investigation will be re-opened as necessary.Device history lot : no anomalies or deviations were found during the review.
|
| |
|
Event or Problem Description
|
|
Medical records reviewed: patient received primary right total hip endoprosthesis on (b)(6) 2007 to address acute femoral head necrosis.On (b)(6) 2020, patient experienced a right hip snapping sound when getting up from a squatting/stooping position, resulting in acute, severe pain in the right hip.When walking, a noticeable squeaking and creaking sound was audible from her hip.There was no related trauma or fall event associated with this onset.Radiological diagnosis showed uncentering of the femoral head, indicative of inlay dislocation with metal-to-ceramic contact of the artificial hip joint."advanced wear is found with decentering of the head and likely complete wear of the plastic inlay.No classic signs of loosening at the cup or shaft were found".This required immediate re-operation with inlay/acetabular replacement.It is noted that this is a single examination.A series of comparative x-rays over time is required to definitively diagnose polyethylene bearing wear by radiographs alone.The revision procedure was conducted on 07/31/2020.Revising surgeon noted that the ceramic femoral head had made contact with the uncovered metal acetabular cup.The radiologist had suggested the possibility of polyethylene bearing wear, but the revising surgeon only noted that the liner was "dislocated and sitting only loosely in the acetabular cup" and was easily removed.There was no mention of any polyethylene bearing wear, just scoring of the cup from the ceramic head contacting it, secondary to the disassociated liner.The radiographic evidence for possible liner wear was not conclusive, only possible, given the current information.If new information becomes available, then this coding mat be updated.Lab results provided are not dated, but with respect to a slightly elevated crp, physician reported: "the blood count and crp results over time were normal.Given normal wound conditions, the continued elevation of crp values is deemed physiological in this postoperative patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : examination of the returned device found evidence to support disassociation of the liner from the cup while implanted, as well as evidence of wear and device fracture.Depuy considers the investigation closed.Should additional info be received, the investigation will be re-opened as necessary.Device history lot :1) quantity manufactured: (b)(4).2) date of manufacture: 30-aug-2007 3) any anomalies or deviations identified in dhr: none 4) expiry date: aug-2012.5) ifu reference: 090200701 corrected: h3.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4) investigation summary: examination of the returned device found evidence to support disassociation of the liner from the cup while implanted, as well as evidence of wear and device fracture.Depuy considers the investigation closed.Should additional info be received, the investigation will be re-opened as necessary.Device history lot: 1) quantity manufactured: (b)(4) 2) date of manufacture: 30-aug-2007.3) any anomalies or deviations identified in dhr: none 4) expiry date: aug-2012 5) ifu reference: (b)(4).
|
| |
|