Model Number PXSLIMLAN115T45
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report numbers: 3005168196-2021-00352, 3005168196-2021-00354.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the internal iliac artery (iia) using a ruby coil, a lantern delivery microcatheter (lantern), pod coils, and a non-penumbra support catheter.During the procedure, the physician advanced a ruby coil into the target vessel using the lantern and support catheter.Then, while the ruby coil started to form in the target vessel, the lantern kicked back into the support catheter and subsequently, the ruby coil, coiled off in the support catheter.Therefore, the physician decided to remove the ruby coil.Upon retraction, the physician experienced resistance and the ruby coil unintentionally detached within the lantern.Therefore, the lantern containing the detached ruby coil was removed.After removing the lantern, the physician noticed that the distal end of the lantern was kinked.Therefore, the lantern was not used for the remainder of the procedure.Afterwards, while attempting to insert a pod coil into a new lantern, the hospital technologist bent the pusher assembly of the pod coil.Therefore, the pod coil was removed.The procedure was completed using a new pod coil, the same second lantern, and a new support catheter.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LANTERN DELIVERY MICROCATHETER
Common Device Name
DQY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key11387858
Report Number3005168196-2021-00353
Device Sequence Number16692034
Product Code DQY
UDI-Device Identifier00814548016627
UDI-Public00814548016627
Combination Product (Y/N)Y
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K152840
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/01/2021
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/20/2023
Device Model NumberPXSLIMLAN115T45
Device Catalogue NumberPXSLIMLAN115T45
Device Lot NumberF99640
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/01/2021
Initial Report FDA Received Date02/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age78 YR
Date Report Sent to FDA01/01/2005
-
-