• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS ARTOURA BREAST TISSUE EXPANDER; EXPANDER, SKIN, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS ARTOURA BREAST TISSUE EXPANDER; EXPANDER, SKIN, INFLATABLE Back to Search Results
Model Number SMXP130RH
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Deformity/ Disfigurement (2360)
Date of Event 12/29/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: deflation.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a female patient underwent a breast reconstruction with a artoura breast tissue expander 475cc and suffered from deflation on the left side.As a result, the patient had undergone removal and replacement with extra hp implants on (b)(6) 2021.
 
Additional Manufacturer Narrative
On 4/26/2021, the mentor failure analysis lab has received the device for evaluation.On 4/28/2021 , during visual evaluation of the tissue expander, the tear was not identified.As a result, leak testing was performed in accordance with mentor procedures, and a tear was identified at the 4 o¿clock position.When evaluated under a microscope the tear pattern did not reveal parallel striations that are consistent with markings made by a sharp instrument.The results also confirmed that there were no abnormalities with the manufacturing of the tissue expander that would have impacted its performance.Therefore, we were unable to identify the root cause of this tear.We would like to assure you that mentor adheres to the highest standards of quality.Our quality assurance process is robust for every product we make and requires that each mentor product undergoes a stringent visual inspection and rigorous testing prior to shipment.In addition, we maintain a comprehensive quality assurance system that complies with the food and drug administration's (fda) good manufacturing practice regulations.Prior to shipping, our devices are checked to ensure that they satisfy these standards in accordance with these processes.We also closely and continually monitor and review the clinical performance and real-world evidence for all our products through clinical studies, registries and post-market surveillance activities.While deflation is historically the most common complication related to tissue expanders, the rate of deflation with mentor tissue expanders remains very low.An incident of this nature may be the result of one or more of the following factors: over inflation of the tissue expander, use of the tissue expander in emerging surgical techniques not identified in the instructions for use (ifu), continuous and sustained stresses to the tissue expander, vigorous body movement (e.G.Physical exercise) or excessive manipulation/ trauma in the region of the expander.At mentor, we are committed to delivering high-quality tissue expanders and an exceptional customer experience.We take your concerns seriously and would like to thank you for taking the time to provide us with the information that was critical for our investigation and our post-market surveillance.We have shared your experience with our research and development team as they consider all customer feedback, and emerging surgical trends as they develop next generation product designs and innovations.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTOURA BREAST TISSUE EXPANDER
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
MDR Report Key11388032
Report Number1645337-2021-02096
Device Sequence Number19654035
Product Code LCJ
UDI-Device Identifier00081317024316
UDI-Public00081317024316
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K161176
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/20/2024
Device Model NumberSMXP130RH
Device Catalogue NumberSMXP130RH
Device Lot Number9448243
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/26/2021
Initial Date Received by Manufacturer 02/09/2021
Supplement Date Received by Manufacturer04/26/2021
Initial Report FDA Received Date02/26/2021
Supplement Report FDA Received Date04/30/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-