|
Device evaluation summary: the product was returned to mentor for evaluation.Mentor conducted a visual inspection of the returned device.Visual analysis of the returned sample determined that the mentor memoryshape¿ tm+ 555cc breast implant was found to be ruptured.In addition, a line of shell wear within the rupture was noted.The evaluation determined that the cause of the rupture is consistent with normal wear.Shell wear suggests in-vivo folding or creasing of the device.This may be the result of the following factors: continuous and sustained stresses to the device such as too small breast pocket and folding or wrinkling of the shell in the breast pocket.In some cases, the breast implants may also wear out over time.It should be noted that as part of mentor¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: rupture.Manufacturer¿s reference number: (b)(4).
|