Model Number 71067322
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/04/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference number: (b)(4).
 
Event or Problem Description
It was reported that during set up inspection process was identified that a gear of the 10mm ratchet wrench was broken.When tightening the gold side did skip a gear tooth inside the ratcheting mechanism.There was no patient involved when this issue was noticed.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
Results of investigation: the device, intended for use in treatment, was not returned for evaluation.Therefore, product analysis could not be performed at this time.So the reported event could not be confirmed.A review of complaint history on the listed part revealed no prior complaints for the listed batch with the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further investigation warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
10MM RATCHET WRENCH
Common Device Name
APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11390985
Report Number1020279-2021-01648
Device Sequence Number14266890
Product Code KTT
UDI-Device Identifier03596010607768
UDI-Public03596010607768
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K093047
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71067322
Device Catalogue Number71067322
Device Lot Number12LCT0030
Initial Date Received by Manufacturer 02/04/2021
Supplement Date Received by Manufacturer03/31/2021
Initial Report FDA Received Date02/28/2021
Supplement Report FDA Received Date04/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-