• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. TENSION GAUGE; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. TENSION GAUGE; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Catalog Number 00578907100
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/02/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the device was found to be broken after procedure.Patient injuries were not reported.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved or product information, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Case-2021-00040251-1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TENSION GAUGE
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11391234
Report Number1020279-2021-01713
Device Sequence Number1166730
Product Code LXH
UDI-Device Identifier00885556618868
UDI-Public885556618868
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number00578907100
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 02/02/2021
Supplement Date Received by Manufacturer03/25/2021
Initial Report FDA Received Date03/01/2021
Supplement Report FDA Received Date03/25/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-