| Model Number |
X SERIES |
| Medical Device Problem Codes |
Reset Problem (3019); Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that while attempting to treat a (b)(6) year-old male patient, the device inappropriately shut down and displayed a "internal error occurred - system reset required" message.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental medwatch report is reporting the evaluation of the device and correcting information submitted on the initial medwatch report.Please reference section b5.Evaluation: the device was returned to zoll medical corporation; the customer's report was observed during review of the device's history log.The device was put through extensive testing including full functional testing and defib function stress testing without duplicating the report.The monitor board was replaced as a precuation.The device passed the final test procedure, was recertified, and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that while attempting to treat a 55-year-old male patient, the device restart/reboot and displayed a "internal error occurred system reset required" message.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
|
| |
|