• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I CERTAIN LOCATOR ABUTMENT 4.1MM(D) X 3MM(H); DENTAL ABUTMENT SCREW

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I CERTAIN LOCATOR ABUTMENT 4.1MM(D) X 3MM(H); DENTAL ABUTMENT SCREW Back to Search Results
Model Number ILOA003
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/02/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zimmer biomet (b)(4).Patient identifier : (b)(6).Weight unknown / not provided.Last name unknown / not provided.
 
Event or Problem Description
It was reported the patient went to the medical practice with the screw of the locator abutment fractured.Fractured portion was removed from the implant.Doctor will place another locator at a later point of time.Locator placed on (b)(6) 2020 and removed on (b)(6) 2021.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
Zimmerbiomet complaint number (b)(4).H3: device evaluated by manufacturer: change ¿no' to 'yes'.The complaint report was received, reviewed, and evaluated by the manufacturer's complaint process.A review of other manufacturer's complaint files and related trending data for similar incidents was performed to evaluate whether other similar complaints have been received and if data suggests any adverse trends may have contributed to the complaint root cause.Condition of the returned complaint related product was evaluated to confirm the reported product complaint.As part of each complaint investigation, the manufacturer performs an analysis to determine the root-cause of the reported problem.Depending upon the information provided, a true root-cause may or may not be determined.In these cases, the most-likely cause of the reported issue would be determined.Other known issues that are inherent to the business operations, typically ones that are related to shipping, handling, customer service, or other, are internally tracked and continually trended for unusual increases.As a result of the above investigation, the manufacturer has concluded the following as the root-cause or most likely root-cause of the reported complaint condition: no manufacturing defect was found.Most likely hazard experience: thread fracture during function.Hazard experience: most likely root cause of the failure mode / hazard: typically is the result from insufficient tightening torque at placement, overloading of the abutment in function, or not retightening the abutment at prescribed intervals.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CERTAIN LOCATOR ABUTMENT 4.1MM(D) X 3MM(H)
Common Device Name
DENTAL ABUTMENT SCREW
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
MDR Report Key11392216
Report Number0001038806-2021-00338
Device Sequence Number16924355
Product Code DZE
UDI-Device Identifier00844868006084
UDI-Public(01)00844868006084(17)250201(10)1235763
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K012911
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 03/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/01/2025
Device Model NumberILOA003
Device Catalogue NumberILOA003
Device Lot Number1235763
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/09/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 02/03/2021
Supplement Date Received by Manufacturer03/05/2021
Initial Report FDA Received Date03/01/2021
Supplement Report FDA Received Date03/24/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age78 YR
-
-