Allergan did not submit this mdr within 30 days of becoming aware.Recent stimulated reporting related to 2011068-7/2/19-001-r has increased complaint and mdr volume.Allergan is implementing a plan to address the increased volumes.A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation: visual analysis of the returned device identified broken shell, crease flat, and missing a piece of shell (0-25%).Weight measurement of the device identified device was underweight.A microscopic analysis was performed which identified a sharp broken edge, non-penetrating nick near the broken site, wear abrasion, and stress marks.A dimensional measurement in the shell was performed which identified the thickness was within specification.Based on the device analysis the final assessment was a sharp broken edge opening on the posterior assessed as surgical impact opening, and missing shell assessed as inconclusive.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|