Allergan did not submit this mdr within 30 days of becoming aware.Recent stimulated reporting related to 2011068-7/2/19-001-r has increased complaint and mdr volume.Allergan is implementing a plan to address the increased volumes.A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation: visual analysis of the returned device identified a curved opening and crease folds.Weight measurement of the device identified device was underweight.A microscopic analysis was performed which identified a curved striated edge opening consistent with use of a surgical tool, wear abrasion, and stress marks.Based on the device analysis the final assessment was a curved striated edge opening on posterior side assessed as surgical damage.The event of "capsular contracture" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade unknown.
|