Model Number VENUM14080
Medical Device Problem Codes Fracture (1260); Material Deformation (2976); Activation Failure (3270)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/03/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history record is currently being performed.The device has not been returned to the manufacturer for evaluation.However, images were provided for review.The investigation of the reported event is currently underway.Expiry date: 11/2022.
 
Event or Problem Description
It was reported that during a stent placement procedure, the stent allegedly got stuck on the catheter upon deployment.It was further reported that the technician ripped out the deployment system causing the stent to fracture.The end of the stent was noted to protrude lateral and was realigned.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: the lot history record of this lot was reviewed with special attention to the manufacturing and inspection of this product, and the product was found to have met the specification prior to shipment.Investigation summary: the physical sample was not available.Two images were provided demonstrating the placed stent inside the vessel.The stent was visibly constricted in the area of the silicone stent brake.Distal and proximal of that section the stent was regularly expanded so that it appeared like an hour-glass which indicated that the deployment sheath was retracted and that the stent adhered to the silicone brake.One image further confirmed strut fracture, elongation, and misalignment as subsequent events which leads to a confirmed result for expansion issue.A definite root cause for the reported event could not be determined.Labeling review: a review of the relevant instructions for use for this product was conducted.The instructions for use was found to address potential complications and adverse events such as incorrect positioning of the stent requiring further stenting or surgery, intimal injury, and malposition.H10: d4 (expiry date: 11/2022).H11: h6 (method, result).H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event or Problem Description
It was reported that during a stent placement procedure, the stent allegedly got stuck on the catheter upon deployment.It was further reported that the technician ripped out the deployment system causing the stent to fracture.The end of the stent was noted to protrude lateral and was realigned.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENOVO VENOUS STENT
Common Device Name
VENOUS STENT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM  76227
MDR Report Key11395166
Report Number9681442-2021-00071
Device Sequence Number11609025
Product Code QAN
UDI-Device Identifier00801741103438
UDI-Public(01)00801741103438
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P180037
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVENUM14080
Device Catalogue NumberVENUM14080
Device Lot NumberANEY3215
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/03/2021
Supplement Date Received by Manufacturer03/08/2021
Initial Report FDA Received Date03/01/2021
Supplement Report FDA Received Date03/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-