• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH 5.5 EXP VERSE SCREW 7.0 X 35; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH 5.5 EXP VERSE SCREW 7.0 X 35; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD Back to Search Results
Model Number 199721735
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/04/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Complainant part is not expected to be returned for manufacturer review/investigation.Reporter is jnj representative.The investigation could not be completed, no product was received; no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2021, the patient underwent a surgery.During removal strips of titanium sheared from the screw head and were spread around the surgical site.The risk to the patient required time to be taken to locate all fragments of the screw thread to ensure none was left in the wound.When removing a 5-level fixation l2-s1 due to infection the correction keys stripped the screws internal thread.This caused small strips of titanium to be spread around the spine.Images of damaged implants and fragments attached.A request has been made for the implants to be sterilized for further inspection.This complaint involves unknown number of devices.This report is for (1) 5.5 exp verse screw 7.0 x 35 this report is 2 of 3 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Customer quality investigation: the implant(s) was not returned, and the investigation will be completed based on the supplied image(s).The image(s) was reviewed, and the complaint condition of broken was confirmed as the image(s) showed the expedium verse drag ring has broken from the tulip head of the screw.As the implant(s) was not returned and as received, dimensional, material or drawing reviews are not applicable.A manufacturing record evaluation could not be performed as the lot number could not be determined from the image.There is no indication that a design or manufacturing issue contributed to the complaint as the circumstances during the time of the event are unknown.No new malfunctions were observed during this investigation (based on the images) that may have contributed to the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device history review: a manufacturing record evaluation could not be performed as the lot number could not be determined from the image.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6- the implant(s) was not returned, and the investigation will be completed based on the supplied image(s) located in pc's attachments section received through 09feb2021.The image(s) was reviewed, and the complaint condition of broken was confirmed as the image(s) showed the expedium verse drag ring has broken from the tulip head of the screw.As the implant(s) was not returned and as received, dimensional, material or drawing reviews are not applicable.A manufacturing record evaluation could not be performed as the lot number could not be determined from the image.There is no indication that a design or manufacturing issue contributed to the complaint as the circumstances during the time of the event are unknown.No new malfunctions were observed during this investigation (based on the images) that may have contributed to the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device history review = a manufacturing record evaluation could not be performed as the lot number could not be determined from the image.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.5 EXP VERSE SCREW 7.0 X 35
Common Device Name
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key11395603
Report Number1526439-2021-00375
Device Sequence Number1165854
Product Code NKB
UDI-Device Identifier10705034442129
UDI-Public10705034442129
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K142185
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number199721735
Device Catalogue Number199721735
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/04/2021
Supplement Date Received by Manufacturer03/03/2021
04/16/2021
Initial Report FDA Received Date03/01/2021
Supplement Report FDA Received Date03/04/2021
04/19/2021
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5.5 EXP VERSE SCREW 7.0 X 35; UNKNOWN LOCKING/SET SCREWS
Outcome Attributed to Adverse Event Required Intervention;
-
-