Patient's exact age is unknown; however, it was reported that the patient was over the age of 18.(b)(4).Investigation results: the returned accutrac 200 laser fiber was analyzed, and a visual evaluation noted that it was returned in one piece.The fiber measured 318.4 cm from the end of the connector to the end of the exposed glass tip.The exposed glass tip measured approximately 4 mm and appeared unused and in good condition.There was distorted light from the sma connector, indicating a fracture within the fiber connector.No other issues with the device were noted.The reported event was confirmed.The analysis of the returned device revealed that there was a distorted light from the sma connector, indicating a fracture within the fiber connector, likely leading to the reported failure to fire.The exposed glass tip appeared unused and in good condition.It is likely that procedural handling of the device during set-up or use contributed to the fracture within the fiber connector.A broken fiber within the connector of the device can result in insufficient aiming beam and low laser energy output, up to a complete inability for the fiber to fire.Review and analysis of all available information indicated that the root cause is adverse event related to procedure.A complaint with a cause of adverse event related to procedure indicates that the adverse event occurred during the procedure and the device had no influence on the event.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
|
|
It was reported to boston scientific corporation on (b)(6) 2021 that an accutrac 200 laser fiber was used in a ureteroscopy procedure in the kidney performed on (b)(6) 2020.Reportedly, the laser unit used was a lumenis 100 watt laser console with the laser settings of 0.5 joules, 40 hertz and 20 watts.It was reported, during procedure, there was no energy traveling through the laser fiber.The procedure was completed with another accutrac 200 laser fiber.There were no patient complications reported as a result of this event.This event has been deemed a reportable event based on the investigation finding of the laser fiber broken within the connector.
|