H3, h6: the device, used in treatment, was not returned for evaluation but the pictures were reviewed, and the loosening was confirmed.The clinical / medical investigation concluded that, no relevant clinical information has been provided to perform a thorough medical investigation.Per the analysis of the provided undated, unlabeled x-ray, the x-ray supports the complaint.However, without the requested clinical information, operative report, revision report, pre/post radiographs the clinical root cause of the reported aseptic loosening cannot be determined.Based on the information provided, the patient was revised, and the tibial and patella components were removed.Therefore, the impact to the patient beyond that which has already been reported cannot be determined.Should any additional relevant patient information be provided, this compliant would be re-assessed.A review of the complaint history for the listed part revealed no prior complaints for the listed failure mode with the same batch number.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Factors and/or some potential probable causes that could contribute to the reported event have been identified as abnormal motion over time, bone degeneration, fit/sizing, lack of ingrowth and/or traumatic injury.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
|