| Model Number |
482677580 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/04/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that two xia serrato medial biased polyaxial screws "popped-out" 5 months post-operatively.Revision has been performed.This report will capture the first of two screws.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual, dimensional, material and functional analysis could not be performed as the device was not returned.Device and complaint history records review could not be performed as a valid lot code was not provided and could not be obtained.Due to a lack of information and the device not being returned, the root cause of the reported event can not be determined.Potential root causes include: user error (unstable construct); incorrect or small-sized screws implanted; poor bone quality of patient; patient experiences a fall or trauma post-surgery; high activity level of patient post-surgery h3 other text : status and location of the device is unknown.
|
| |
|
Event or Problem Description
|
|
It was reported that two xia serrato medial biased polyaxial screws "popped-out" 5 months post-operatively.Revision has been performed.This report will capture the first of two screws.
|
| |
|