| Model Number |
MC2-5090S |
| Medical Device Problem Code |
Migration or Expulsion of Device (1395)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Even though no product has been returned, fsca investigation was performed and variation in the torque applied to the threaded cap during the assembly process was identified as the root cause of the failure.Per the manufacturing instructions, the threaded cap must be tightened to 40 in-lbs.Although all operators followed the assembly procedure, and the torque wrench indicated 40 in-lbs was applied, the manner of using the torque wrench resulted in variances in applied torque.The manner in which the operator handles the torque wrench may impact the actual torque applied to the end cap, thereby creating a false positive that the specified torque has been applied.If the specified torque is not applied, the effectiveness of the cap tightening process may be compromised.Device returned to lirc.
|
| |
|
Event or Problem Description
|
|
Information was received that a revision procedure was performed on (b)(6) 2020 due to the end cap being separated.During a review of investigation findings from lirc it was identified that there was no problem found.
|
| |
|
Event or Problem Description
|
|
During a review of investigation findings from lirc it was identified that the end-cap was separated.
|
| |
|