BOSTON SCIENTIFIC CORPORATION MUSTANG; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number 24674
Medical Device Problem Codes Break (1069); Difficult to Remove (1528); Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/03/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that balloon rupture, shaft break, and difficulty removal occurred.The completely occluded target lesion was located in the non-tortuous and non-calcified right subclavian vein with stent occlusion.A 12.0 x 80, 75cm mustang balloon catheter was advanced for dilation.However, on initial inflation at 12 atmospheres for 10-15 seconds, the balloon ruptured and was completely torn in half.The peripheral half of the balloon bunched and could not be withdrawn from the sheath.A snare was advanced over the balloon and attempted to remove it through the access.The procedure was completed with a non-boston scientific balloon catheter.There were no patient complications nor injuries reported and the patient was doing well.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MUSTANG
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key11397906
Report Number2134265-2021-02472
Device Sequence Number3381833
Product Code FGE
UDI-Device Identifier08714729794745
UDI-Public08714729794745
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K141521
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/16/2023
Device Model Number24674
Device Catalogue Number24674
Device Lot Number0026528066
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/04/2021
Initial Report FDA Received Date03/02/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-