| Model Number |
0279251101 |
| Medical Device Problem Code |
Excessive Heating (4030)
|
| Health Effect - Clinical Codes |
Burn, Thermal (2530); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/29/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient was burned and the device was continuously activating.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient was burned and the device was continuously activating.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: burn injury.Probable root cause: heat can result from insufficient suction which can result from: inadequate hospital suction, leak, bad connection to hospital suction line clogging caused by suction path blockage, lower electrode mass due to variation in electrode thickness or opening cut.Use error: the product was not returned for investigation and the reported failure mode was not confirmed.The failure mode will be monitored for future reoccurrence.H3 other text: 81.
|
| |
|