U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC CANN SCREW MEASURING DEVICE FOR 6.5MM & 7.3MM CANN SCREWS; GAUGE, DEPTH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC CANN SCREW MEASURING DEVICE FOR 6.5MM & 7.3MM CANN SCREWS; GAUGE, DEPTH Back to Search Results
Model Number 319.70
Medical Device Problem Code Incorrect Measurement (1383)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/08/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient identifier: (b)(6).Complainant part is not expected to be returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2021, 2 cannulated screws were put into a patient¿s ankle and were be too long.When two depth gauges were compared, it was determined that the depth gauge in question, was off at least 5mm.The screws were exchanged, and surgery successfully completed.There were no fragments generated.The surgery was delayed for 5 minutes.Concomitant devices reported: 6.5mm cannulated screw fully threaded 75mm (part number 208.471, lot unknown, quantity 1).6.5mm cannulated screw fully threaded 80mm (part number 208.472, lot unknown, quantity 1).This report involves 1 cann screw measuring device for 6.5mm & 7.3mm cann screws.This is report 1 of 3 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Customer quality investigation: the instrument(s) was not returned, and the investigation will be completed based on the supplied image(s).The image(s) was reviewed, and there is some nicks on the instrument, however it would not affect the functionality of the device.There is no indication the measurements of the depth gauge are off as the comparison of the two gauges are not perfectly lined up to make an accurate conclusion and therefore is unconfirmed.As the instruments) was not returned an as received, dimensional, material or drawing reviews are not applicable.A manufacturing record evaluation could not be performed as part/lot combination are unknown at synthes gmbh, no dhr review possible.There is no indication that a design or manufacturing issue contributed to the complaint as it was unconfirmed.No new malfunctions were observed during this investigation (based on the images) that may have contributed to the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device history lot: part/lot combination are unknown at synthes gmbh, no dhr review possible.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Visual inspection: cann scr measure dv 6.5 & 7.3 cann scrs (part# 319.70, lot# a4df803, qty# 1) was returned and received at us customer quality (cq).Upon visual inspection at cq, it is observed that the device looks good without any physical damage.The surface of the device shows normal wear consistent with the device use which would not contribute to the complaint condition.Document/specification review the following document(s) was reviewed: -cann screw meas no design issues or discrepancies were found during this investigation.Dimensional inspection (calipers: ca472): dimensional inspection of the received device was performed at customer quality (cq).The length of the device is within specification as per the drawing.Complaint confirmed? no investigation conclusion: the complaint cannot be confirmed for cann scr measure dv 6.5 & 7.3 cann scrs (part# 319.70, lot# a4df803) as no defects were identified with the returned device.A definitive root cause could not be identified for the reported problem.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/specification review.Based upon these results, no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CANN SCREW MEASURING DEVICE FOR 6.5MM & 7.3MM CANN SCREWS
Common Device Name
GAUGE, DEPTH
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key11399883
Report Number2939274-2021-01124
Device Sequence Number11054328
Product Code HTJ
UDI-Device Identifier10886982190314
UDI-Public(01)10886982190314
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number319.70
Device Catalogue Number319.70
Device Lot NumberA4DF803
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/22/2021
Initial Date Received by Manufacturer 02/08/2021
Supplement Date Received by Manufacturer03/09/2021
03/22/2021
Initial Report FDA Received Date03/02/2021
Supplement Report FDA Received Date03/10/2021
04/07/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
6.5MM CANNULATED SCREW FULLY THREADED 75MM; 6.5MM CANNULATED SCREW FULLY THREADED 80MM
Patient Age36 YR